Thursday, January 31, 2008
Comment p5000etch SNF 2008-01-31 14:37:19: Ch.C down for handling.
Intermittently faults for encoder error when running in Auto but not in manual.
Re: Shutdown p5000etch SNF 2008-01-31 04:02:14: wafer stuck in Ch C
Robot was stuck on the door. Straightened out the robot and unloaded user's wafer. Now intermittently faulting for an extension encoder error when running ch.C in auto. Does not fail when moving wafer in and out the chamber in manual. will continue to troubleshoot tomorrow. Cycled 30 wafers though Ch.B with no problems.
Wednesday, January 30, 2008
Comment p5000etch SNF 2008-01-30 15:14:14: Ch.A down for heat exchanger
Heat exchange shuts down because of a flow fault. Possibly a defective flow meter. Need to troubleshoot further.
Re: Problem p5000etch SNF 2008-01-28 18:42:35: Trouble with robot arm
Cleaned and reseated two connectors in the robot drive assembly. Changed system constant 107 (comb sensor test status) from 2 to 1. When it was set at 2, the system kept failing for "comb sensor not in correct location". Set at 1, error has not reoccured after 68 wafers cycled through Ch.C.
Re: Shutdown p5000etch SNF 2008-01-29 14:42:51: Robot extension problem
Cleaned and reseated two connectors in the robot drive assembly. Changed system constant 107 (comb sensor test status) from 2 to 1. When it was set at 2, the system kept failing for "comb sensor not in correct location". Set at 1, error has not reoccured after 68 wafers cycled through Ch.C. Ran "Ch.C Hanjun" recipe to verify chamber operations.
Comment p5000etch SNF 2008-01-30 12:37:49: Update
Cleaned and reseated two connectors in the robot drive assembly. Changed system constant 107 (comb sensor test status) from 2 to 1. When it was set at 2, the system kept failing for "comb sensor not in correct location". Set at 1, error has not reoccured after 20 wafers cycled through Ch.C. Will cycle wafers until 3 pm. Will release system if no other faults.
Tuesday, January 29, 2008
Shutdown p5000etch SNF 2008-01-29 14:42:51: Robot extension problem
Robot extension errors are occuring more frequently now. Power supply voltages checked ok. Possible back robot cable. Will continue to troubleshoot.
Re: Problem p5000etch SNF 2008-01-29 06:37:59: One wafer is stuck in load lock chamber
Removed user's stuck wafer.
Shutdown p5000etch SNF 2008-01-29 08:08:30: wafer is stuck in LL
By looking inside LL, wafer falls down from arm
Problem p5000etch SNF 2008-01-29 06:37:59: One wafer is stuck in load lock chamber
When I started giving one wafer into LL,
Robot was stuck when it picked up the wafer.
I followed the procedure in the manual and robot returned
back to LL. However, it seems robot lost its position and
couldn't go to home position.
The errors which I saw are:
(1)Encorder error
(2)Cannot go to elevator slot with wafer
(3)Cannot return wafer on vacuum chack until robot is homed
(4)Storage position is not known.
Robot was stuck when it picked up the wafer.
I followed the procedure in the manual and robot returned
back to LL. However, it seems robot lost its position and
couldn't go to home position.
The errors which I saw are:
(1)Encorder error
(2)Cannot go to elevator slot with wafer
(3)Cannot return wafer on vacuum chack until robot is homed
(4)Storage position is not known.
Monday, January 28, 2008
Problem p5000etch SNF 2008-01-28 18:42:35: Trouble with robot arm
I ran 12 dummy wafers in 3 lots (with 4 wafers per lot) in Chamber C. The outcome was:
1) First lot etched fine, but gave an error "Cannot complete: Robot extension position is not known".
2) Second lot didn't enter the etch chamber C at all. The same error as above showed up before the wafers could be loaded into the chamber.
3) Third lot proceeded fine upto 2 wafers and then the third wafer was not loaded into the chamber C because of the same error as above.
Upon checking the error log, it appears as though the encoder on the robot arm motor has problems.
1) First lot etched fine, but gave an error "Cannot complete: Robot extension position is not known".
2) Second lot didn't enter the etch chamber C at all. The same error as above showed up before the wafers could be loaded into the chamber.
3) Third lot proceeded fine upto 2 wafers and then the third wafer was not loaded into the chamber C because of the same error as above.
Upon checking the error log, it appears as though the encoder on the robot arm motor has problems.
Comment p5000etch SNF 2008-01-28 16:23:47: Side B works
I used Ch C, used side B to load the wafers, it worked alright.
Re: Problem p5000etch SNF 2008-01-26 20:34:57: Ch.C loading/unloading
Adjusted the slit door opening speed. Cycled 8 wafers through the chamber without problems.
Re: Problem p5000etch SNF 2008-01-26 21:05:13: Wafer stuck inside
Adjusted the slit door opening speed. Cycled 8 wafers through the chamber without problems.
Re: Problem p5000etch SNF 2008-01-27 11:32:28: Ch C down
Adjusted the slit door opening speed. Cycled 8 wafers through the chamber without problems.
Sunday, January 27, 2008
Problem p5000etch SNF 2008-01-27 11:32:28: Ch C down
I managed to recover the system from the frozen state and put Ch C offline. Note that the previous user's wafer is still stuck in Chamber C.
You can use other chambers, but I recommend using it in the manual mode. It has been giving a lot of errors related to loading and unloading.
You can use other chambers, but I recommend using it in the manual mode. It has been giving a lot of errors related to loading and unloading.
Saturday, January 26, 2008
Problem p5000etch SNF 2008-01-26 21:05:13: Wafer stuck inside
One of my wafers got stuck in the chamber C. There was an error stating "Chamber C slit valve did not open in maximum allowed time". After waiting a little, I tried the recovery procedure...but could not make much progress as I could not switch to manual mode. Fortunately, it is a dummy wafer.
Problem p5000etch SNF 2008-01-26 20:34:57: Ch.C loading/unloading
Door slit fails to close/open error message still occurs occasionally.
Friday, January 25, 2008
Problem p5000etch SNF 2008-01-25 15:18:39: use load side B...
side A, when unloading, pulls the wafer out ~1/2 inch
May be too close to LL door & might smash your wafer
May be too close to LL door & might smash your wafer
Re: Problem p5000etch SNF 2008-01-09 09:12:02: CHB still down. CHA/C OK to use.
Installed rebuilded turbo pump and controller and also,
changed change clamping kit and pedestal quartz ring.
Chamber will need to be condition before running sample.
changed change clamping kit and pedestal quartz ring.
Chamber will need to be condition before running sample.
Re: Problem p5000etch SNF 2008-01-24 11:37:40: Cassette A clamp not opening
Repaired cassette clamp.
Thursday, January 24, 2008
Problem p5000etch SNF 2008-01-24 11:37:40: Cassette A clamp not opening
Cassette A clamp is not fully opening. Tried adjusting the pneumatic actuator but still not working properly.
Please use Cassette B.
Please use Cassette B.
Re: Shutdown p5000etch SNF 2008-01-23 20:54:47: Wafer stuck while unloading
Adjusted Ch.C slit door pneumatic actuator speed. Cycled many times with no problems. Moved wafer in and out of the chamber with no problems.
Wednesday, January 23, 2008
Shutdown p5000etch SNF 2008-01-23 20:54:47: Wafer stuck while unloading
Error message: Chamber C door slit did not close within time limit
Tuesday, January 22, 2008
Re: Shutdown p5000etch SNF 2008-01-17 08:34:40: Heatexchanger 0 has bo flow.
heatexchanger is back up
Thursday, January 17, 2008
Wednesday, January 16, 2008
Problem p5000etch SNF 2008-01-16 20:53:30: both HX's dead.
HX1 and HX2 appear to have tripped off. Indicating
low fluid, and temp out of setpoint.
low fluid, and temp out of setpoint.
Shutdown p5000etch SNF 2008-01-16 14:29:28: Wafers stuck
Was running Ch.C Poly Etch
Error: Chamber C door slit did not close within time limit
Error: Chamber C door slit did not close within time limit
Tuesday, January 15, 2008
Problem p5000etch SNF 2008-01-15 16:55:45: Why is chamber C offline?
I tried to use Ch. C for Si etch but found it to be offline. Is there any reason why Ch C is offline? I could not find any information on Coral. Thanks
Monday, January 14, 2008
Comment p5000etch SNF 2008-01-14 12:18:20: Turbo pump
The turbo pump bearing for chambert b seized up, I'm
looking for a replacement. I will update as soon as I find
a replacement turbo pump.
looking for a replacement. I will update as soon as I find
a replacement turbo pump.
Friday, January 11, 2008
Re: Shutdown p5000etch SNF 2008-01-10 15:15:43: wafer broke while unloading
No problems with the handler. User logged problem on the wrong tool down.
Thursday, January 10, 2008
Shutdown p5000etch SNF 2008-01-10 15:15:43: wafer broke while unloading
Unloading arm broke the wafer.
Wednesday, January 9, 2008
Tuesday, January 8, 2008
Comment p5000etch SNF 2008-01-08 10:12:52: Turbo controller for Cha. B
The rebuild turbo controller that I order before the shutdown
has a problem with the eeprom, I called the vendor where
I ordered the controller from and they will send another
controller asap.
has a problem with the eeprom, I called the vendor where
I ordered the controller from and they will send another
controller asap.
Thursday, January 3, 2008
Problem p5000etch SNF 2008-01-03 17:37:50: Ch B offline for pump repair.
Ch A and Ch C are ok to use.
Subscribe to: