P5000 Email Log
Sunday, November 1, 2009
Re: Problem p5000etch SNF 2009-11-01 11:31:23: ChB Backside He
Vented chamber again and noticed one of the new small
lip seal and has a cracked on the edge. Changed small
lip seal and verified b/s he leak at 0.3sccm.
No comments:
Post a Comment
‹
›
Home
View web version
No comments:
Post a Comment