Just two reminders:
1. Check acid change frequency and to log your acid changes. According to the wbdiff operating procedures:
- The 4:1 sulfuric/peroxide clean should be changed once every five days.
- The 5:1:1 H2O:H2O2:HCl hot pot should be changed every four hours.
- The 50:1 HF tank should be changed once/day.
- The 6:1 BOE is changed once per month or on an as-needed basis.
2. Ask before sharing a wet bench. The person who has the bench enabled is responsible for the bench, and therefore, your use if you share. The person enabled has priority access -- and may ask you to enable over him/her.
Thanks for your attention --
Your SNF staff
-- Mary X. Tang, Ph.D. Stanford Nanofabrication Facility CIS Room 136, Mail Code 4070 Stanford, CA 94305 (650)723-9980 mtang@stanford.edu http://snf.stanford.edu